OWLTM Overview : Hardware
Introduction
This document describes the hardware components of the OWLTM system. It is written to give an overview of the system on the highest level. It shows the components, connections between the components, connections between the OWLTM and it's platform (ship/buoy/rig) and specifications.
Detailed datasheet OWLTM Shipborne and OWL MAPTM is attached.
Description of the measurement process is described in introduction to measurement physics?.
Block Diagram
The system has two components:
- Junction box that holds circuit breakers, power-line communication modem and server. The junction box needs a LAN cable and a regular AC current from mains.
- Measurement device enclosed in a high durability, climate controlled aluminium box.
The two components are connected via regular two leaded electrical wire.
Specifications
Operational | |
---|---|
Sensing | 5 to 30 m |
Conditions of operation | Continuous, Day and Night |
Sensitivity for oil products: | |
Min concentration in water column | 1ppm |
Min oil thickness on water surface | 1μm |
Maximum detection depth | 3m |
Pulse Repetition (Sampling) rate | Up to 100Hz |
Sensing laser | Excimer type, XeCl, 308nm |
Hyperspectral detector | Ultraviolet(UV) and visible spectral range |
GPS | Integrated |
Control and communication | |
Operational control | Integrated micro-controller |
Data storage | Local and central servers |
Setup of device operation | Remote, via local server, embedded WEB |
Communication line | Power-line communication |
Alarm processing | Automatic |
Data visualization on the map | Real-Time |
General | |
Power consumption | 150 W/(1500 W arctic) |
Input voltage | 230VAC/50/60Hz |
Dimensions (L x W x H) | 65 cm x 45 cm x 37 cm |
Weight | 40 kg |
List of identifiable oil types
- Light oil
- Medium oil
- Crude oil
Safety The LASER in the measurement device is eye safe. See declaration of safety.
System installation
The Measurement device has robust attachment points along the longest edges. The customers needs to specify where the device is meant to be installed. A drawing of the railing is necessary for us to manufacture the appropriate fastenings.
The Junction Box should be installed indoors. The enclosure of the Junction Box is not weatherproof nor climate controlled. It is a metal box that will be mounted on a wall. It requires power and Local Area Network connections. Instruction for
Example of installed measurement device is shown below.
Attachments (7)
-
OV logo neutral medium.jpg
(67.7 KB) -
added by indrek 9 years ago.
Company logo
- Schematic of OWL.png (14.3 KB) - added by indrek 9 years ago.
- Photo of Measurement Device.png (378.6 KB) - added by indrek 9 years ago.
- Safety_sheet_B22.pdf (106.8 KB) - added by indrek 9 years ago.
- Photio of junction box.png (1.6 MB) - added by indrek 9 years ago.
- Installed measurement device.png (191.1 KB) - added by indrek 9 years ago.
- Ocean_Visuals_OWL_datasheet.pdf (2.0 MB) - added by indrek 9 years ago.
Download all attachments as: .zip