    33= Gas Refill Instructions 
    44 
    5 == Instructions 
     5== 1.-3. Pre-checks 
    66  1. Open a browser an enter the IP address of the LIDAR you wish to monitor. 
    77     You'll also need the username and password for accessing the [wiki:LidarHttpServer LIDAR's web interface].  
     
    1010     Find specific menus, controls and boxes which are described in this manual and make familiar with them.'''  
    1111 
    12 === 4. Check for Laser pulses from last gas refill 
     12== 4. Check for Laser pulses from last gas refill 
    1313  1. From //System Menu// select //Monitor & Control// 
    1414  2. At the bottom of page select monitorlog.txt from Internal storage box [#fig-14 (Fig. 14)] 
     
    2121     Lidar work mode box in //System Setting// [#fig-1 (Fig 1.)] 
    2222 
    23 === 5. Preparation for gas refill 
     23== 5. Preparation for gas refill 
    2424  1. From //System Menu// select //System Settings// [#fig-1 (Fig. 1)] 
    2525  2. In Lidar work mode box [#fig-4 (Fig 4)] you will see 4 modes of operation 
     
    3232     must be below 45(50)^([#note-1 1])^. Status messages: all are Ready and Cyl. Closed. 
    3333 
    34 === 6. Starting gas refill 
     34== 6. Starting gas refill 
    3535  1. From //System Menu// select //System Settings// 
    3636  2. In Lidar work mode box select //Refill laser*//, press //Change//. 
     
    3838  3. From //System Menu// select //Monitor & Control// 
    3939 
    40 ==== Following gas refill 
     40=== Following gas refill 
    4141Now follow the information from Laser and Gas refill system boxes [#fig-4 (Fig. 4)] to [#fig-13 (Fig. 13)] 
    4242  1. __Message: On start__ shows 30 seconds countdown timer 
     
    6161     The pressure values in laser and cylinder will be around 650kPa and 500kPa. 
    6262 
    63 === 7. Finishing the gas refill 
     63== 7. Finishing the gas refill 
    6464  1. Check the pressure values of Laser and Cylinder^([#note-2 2])^ Normal rages are around 650kPa and 500kPa. 
    6565  2. From //System Menu// select //System Settings// 
     
    7171     An example of water spectrum with Raman scattering line is shown in [#fig-16 (Fig. 16)]. 
    7272 
    73 === 8. Troubleshooting 
     73== 8. Troubleshooting 
    7474In any case of emergency STOP lidar operation and contact with our service team 
    7575using is[at]ldi-innovation[dot]eu email. Try to provide as many information as