| 1 | | = LIDAR Gas Refill Instructions |
| | 1 | = Gas Refill Instructions |
| | 2 | |
| | 3 | == Figures |
| | 4 | === Figure 1 - System Settings #fig-1 |
| | 5 | [[Image(01-gas-refill-system-settings.png, nolink)]] |
| | 6 | |
| | 7 | === Figure 2 - Monitor & Control #fig-2 |
| | 8 | [[Image(02-gas-refill-monitor-control.png, nolink)]] |
| | 9 | |
| | 10 | === Figure 3 - Refill Laser Procedure is Selected #fig-3 |
| | 11 | [[Image(03-gas-refill-laser-procedure.png, nolink)]] |
| | 12 | |
| | 13 | === Figure 4 - Laser, Gas Refill Box #fig-4 |
| | 14 | Laser box: |
| | 15 | * Temperature - the first number shows lidar internal ambient air temperature and the number in brackets () shows the temperature of electronics, |
| | 16 | * Energy (uJ) – laser pulse energy, Pressure (kPa) – laser gas pressure. |
| | 17 | |
| | 18 | Gas Refill System box: |
| | 19 | * Cylinder (kPa) – internal fresh gas cylinder pressure before the pressure regulator |
| | 20 | * Lidar (kPa) – internal ambient pressure of lidar hull |
| | 21 | |
| | 22 | [[Image(04-gas-refill-temp-pressure.png, nolink)]] |
| | 23 | |
| | 24 | === Figure 5 - Gas refill system pipes are cleaned. Preparation for laser chamber pumping #fig-5 |
| | 25 | [[Image(05-gas-refill-pipes-cleaned.png, nolink)]] |
| | 26 | |
| | 27 | === Figure 6 - Old gas is pumped out from laser #fig-6 |
| | 28 | [[Image(06-gas-refill-old-gas-removal.png, nolink)]] |
| | 29 | |
| | 30 | === Figure 7 - Old gas is pumped out from laser #fig-7 |
| | 31 | [[Image(07-gas-refill-old-gas-removal.png, nolink)]] |
| | 32 | |
| | 33 | === Figure 8 - Cylinder is working. Laser chamber pressure is almost 0 #fig-8 |
| | 34 | [[Image(08-gas-refill-cylinder-working.png, nolink)]] |
| | 35 | |
| | 36 | === Figure 9 - Cylinder is opened #fig-9 |
| | 37 | [[Image(09-gas-refill-cylinder-open.png, nolink)]] |
| | 38 | |
| | 39 | === Figure 10 - Laser chamber is filled with fresh gas from cylinder #fig-10 |
| | 40 | [[Image(10-gas-refill-fresh-gas.png, nolink)]] |
| | 41 | |
| | 42 | === Figure 11 - Gas cylinder is closing #fig-11 |
| | 43 | [[Image(11-gas-refill-cylinder-close.png, nolink)]] |
| | 44 | |
| | 45 | === Figure 12 - Pressure is stablilized before and after the gas cylinder pressure regulator #fig-12 |
| | 46 | [[Image(12-gas-refill-pressure-stable.png, nolink)]] |
| | 47 | |
| | 48 | === Figure 13 - Gas refill finished#fig-13 |
| | 49 | [[Image(13-gas-refill-finished.png, nolink)]] |
| | 50 | |
| | 51 | === Figure 14 - Internal Storage, monitorlog.txt #fig-14 |
| | 52 | [[Image(14-gas-refill-internal-storage.png, nolink)]] |
| | 53 | |
| | 54 | === Figure 15 - Time From Last Gas refill in seconds and LaSeR Pulses From Last Gas refill #fig-15 |
| | 55 | [[Image(15-gas-refill-last-refill.png, nolink)]] |
| | 56 | |
| | 57 | === Figure 16 - Clean Water Signal Example #fig-16 |
| | 58 | Peak on the left is Raman scattering signal. Broad shape on the right is Dissolved Organic Matter (DOM) fluorescence. |
| | 59 | |
| | 60 | [[Image(16-gas-refill-clean-water-signal.png, nolink)]] |
| | 61 | |